Author:
Zhao Meihong,Jiang Yanxiu,Yang Shuo,Li Wenhao
Abstract
A third-order aberration theory has been developed for the Offner imaging spectrometer comprising an extended source; two concave mirrors; a convex diffraction grating; and an image plane. Analytic formulas of the spot diagram are derived for tracing rays through the system based on Fermat’s principle. The proposed theory can be used to discuss in detail individual aberrations of the system such as coma, spherical aberration and astigmatism, and distortion together with the focal conditions. It has been critically evaluated as well in a comparison with exact ray tracing constructed using the commercial software ZEMAX. In regard to the analytic formulas, the results show a high degree of practicality.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Reference26 articles.
1. Optical system design for a shortwave infrared imaging spectrometer;Huang,2012
2. Freeform spectrometer enabling increased compactness;Jacob;Light Sci. Appl.,2017
3. Optical Design and Tolerancing of a Hyperspectral Imaging Spectrometer;Liu,2016
4. Compact high-resolution spectrometer using two plane gratings with triple dispersion
5. Analytical design of an Offner imaging spectrometer
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献