Cable-driven parallel robots are suitable candidates for rehabilitation due to their intrinsic flexibility and adaptability, especially considering the safety of human–robot interaction. However, there are still some challenges to apply cable-driven parallel robots to rehabilitation, one of which is the geometric calibration. This paper proposes a new automatic calibration method that is applicable for cable-driven parallel rehabilitation robots. The key point of this method is to establish the mapping between the unknown parameters to be calibrated and the parameters that could be measured by the inner sensors and then use least squares algorithm to find the solutions. Specifically, the unknown parameters herein are the coordinates of the attachment points, and the measured parameters are the lengths of the redundant cables. Simulations are performed on a 3-DOF parallel robot driven by four cables for validation. Results show that the proposed calibration method could precisely find the real coordinate values of the attachment points, with errors less than 10 − 12 mm. Trajectory simulations also indicate that the positioning accuracy of the cable-driven parallel robot (CDPR) could be greatly improved after calibration using the proposed method.