Abstract
Paralytic shellfish toxins (PSTs) are a large group of biotoxins that cause paralytic shellfish poisoning. Their appearance in natural waters and their ingestion by aquatic species have a huge socio-economic impact, whereby their monitoring is of the upmost relevance to minimize the consequences. For earlier detection and faster response/action by stakeholders, validation of adjusted analytical methods, particularly for lower concentration levels, is important. This work proposes a derived High-Performance Liquid Chromatography method, with fluorescence detection (HPLC-FLD). The main differences from the official method are the size of the HPLC column and the gradient elution conditions. It covers the current eleven certified reference materials (CRM) available on the market, including the most recent—C3,4. This first calibration report for C3,4 suggests limits of detection (LOD) and limits of quantification (LOQ) of 6 nM and 19 nM (~5 µg STX.2HCl eqv./kg and 17 µg STX.2HCl eqv./kg), respectively. For the remaining CRM, LODs ranged between 3 and 28 nM (~0.9 and 127 µg STX.2HCl eqv./kg), while LOQs varied between 11 and 94 nM (~3 and 409 µg STX.2HCl eqv./kg, considering toxicity equivalency factors (TEFs) reported by EFSA).
Subject
Health, Toxicology and Mutagenesis,Toxicology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献