Author:
Chen Ming,Zhang Miao,Wang Xuming,Yang Qingliang,Wang Maohua,Liu Gang,Yao Lan
Abstract
In this paper, an all-solid-state nitrate doped polypyrrole (PPy(NO3−) ion-selective electrode (ISE) was prepared with a nanohybrid composite film of gold nanoparticles (AuNPs) and electrochemically reduced graphene oxide (ERGO). Preliminary tests on the ISE based in-situ soil nitrate–nitrogen (NO3−-N) monitoring was conducted in a laboratory 3-stage column. Comparisons were made between the NO3−-N content of in-situ soil percolate solution and laboratory-prepared extract solution. Possible influential factors of sample depth, NO3−-N content, soil texture, and moisture were varied. Field-emission scanning electron microscopy (FESEM) and X-ray powder diffraction (XRD) characterized morphology and content information of the composite film of ERGO/AuNPs. Due to the performance excellence for conductivity, stability, and hydrophobicity, the ISE with ERGO/AuNPs illustrates an acceptable detection range from 10−1 to 10−5 M. The response time was determined to be about 10 s. The lifetime was 65 days, which revealed great potential for the implementation of the ERGO/AuNPs mediated ISE for in-situ NO3−-N monitoring. In-situ NO3−-N testing results conducted by the all-solid-state ISE followed a similar trend with the standard UV-VIS method.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
32 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献