Setup Optimization in Ocular Proton Therapy at the National Centre for Oncological Hadrontherapy: Comparison of Two Approaches to Refine the Position of an Eye-Tracking Device

Author:

Sellaro Giulia1,Pella Andrea1,Pepa Matteo1ORCID,Galante Federica1,Ciocca Mario2,Fiore Maria Rosaria3,Chalaszczyk Agnieszka3ORCID,Paganelli Chiara4,Rotondi Marco3ORCID,Vai Alessandro2,Orlandi Ester3ORCID,Baroni Guido14

Affiliation:

1. Bioengineering Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

2. Medical Physics Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

3. Radiotherapy Unit, Clinical Department, National Centre for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

4. Department of Electronics, Information and Bioengineering, Politecnico di Milano (POLIMI), 20133 Milan, Italy

Abstract

This study describes a method for setup optimization in patient simulation for ocular proton therapy (OPT) at the National Center for Oncological Hadrontherapy (CNAO) in Pavia, Italy, with the aim of minimizing the occupancy time of clinical areas and streamlining the actual procedure. Setup repeatability is ensured by patient-specific immobilization tools and relies on the patient’s ability to maintain a stable gaze direction according to the treatment plan. This is facilitated by aligning a light source (LED) on a patient-specific base along the prescribed gaze direction. At CNAO, a dedicated Eye-Tracking System (ETS) was designed to provide the patient with a visible source of light aligned to the desired gaze direction. The ETS position is defined prior to treatment planning, relying on optical-tracking guidance and comparing the position of passive markers fixed on the ETS chassis with patient-specific models prepared offline in accordance with the desired geometry. OPT at CNAO started in 2016 and may be considered as a consolidated clinical routine. However, all the preparation phases, including patient-specific ETS models and setup, still require long sessions in clinical areas such as the computed tomography (CT) and the treatment rooms, with a non-negligible impact on other activities. This study describes a novel approach for patient-specific definition of the ETS position and orientation, aiming at minimizing the time required for preparatory activities inside clinical areas. To minimize the occurrence of biases and to reproduce as much as possible a real end-to-end approach, we included in the analysis data of patients that received OPT in our facility. The study was performed in parallel, carrying out the alignment with the standard method currently used in the clinical workflow of CNAO and with the proposed method. Results are presented as 3D residuals and gaze deviations, comparing ETS alignment based on the new approach with respect to the clinical standard method. The preliminary results of this study are evidence of the capability of the procedure to align the ETS position, allowing performing of the procedure in a non-clinical dedicated room.

Publisher

MDPI AG

Reference27 articles.

1. Practice considerations for proton beam radiation therapy of uveal melanoma during the coronavirus disease pandemic: Particle therapy co-operative group ocular experience;Mishra;Adv. Radiat. Oncol.,2020

2. Proton therapy for the management of uveal melanoma and other ocular tumors;Mishra;Chin. Clin. Oncol.,2016

3. Radiation therapy for choroidal melanoma;Finger;Surv. Ophthalmol.,1997

4. Practice Patterns Analysis of Ocular Proton Therapy Centers: The International OPTIC Survey;Hrbacek;Endocrine,2016

5. Results of proton therapy of uveal melanomas treated in Nice;Courdi;Endocrine,1999

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3