Abstract
The paper presents the validation of an electrochemical procedure for on-site Hg2+ ions determination in wastewater samples using a modified carbon screen-printed electrode (SPE) with a complexing polymeric film based on poly(2,2′-(ethane-1,2-diylbis((2-(azulen-2-ylamino)-2-oxoethyl)azanediyl))diacetic acid) (polyL). Using metal ions accumulation in an open circuit followed by anodic stripping voltammetry, the SPE-polyL electrode presents a linear range in the range of 20 µg/L to 150 µg/L, with a limit of detection (LOD) = 6 µg/L, limit of quantification (LOQ) = 20 µg/L, and an average measurement uncertainty of 26% of mercury ions. The results obtained in situ and in the laboratory using the SPE-polyL modified electrode were compared with those obtained by the atomic absorption spectrometry coupled with the cold vapor generation standardized method, with the average values indicating excellent recovery yields.