Abstract
To conduct burst-echo imaging with air-coupled capacitive micromachined ultrasonic transducers (CMUTs) using the same elements in transmission and reception, this work proposes a dedicated and integrated front-end circuit board design to build an imaging system. To the best of the authors’ knowledge, this is the first air-coupled CMUT burst-echo imaging using the same elements in transmission and reception. The reported front-end circuit board, controlled by field programmable gate array (FPGA), consisted of four parts: an on-board pulser, a bias-tee, a T/R switch and an amplifier. Working with our 217 kHz 16-element air-coupled CMUT array under 100 V DC bias, the front-end circuit board and imaging system could achieve 22.94 dB signal-to-noise ratio (SNR) in burst-echo imaging in air, which could represent the surface morphology and the three-dimensional form factor of the target. In addition, the burst-echo imaging range of our air-coupled CMUT imaging system, which could work between 52 and 273 mm, was discussed. This work suggests good potential for ultrasound imaging and gesture recognition applications.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry