Abstract
We have developed a transmission-based, open-ended coaxial dielectric probe that can be used in clinical situations and overcomes many of the limitations related to the typical reflection-based dielectric probes. The approach utilizes the low profile, open-ended coaxial cables enabling clinicians to still probe relatively compact spaces. The sensing depth can be extended to as large as 1.5 to 2 cm compared with the more typical range of 0.3 mm for conventional probes and is dramatically less affected by measurement technique variability including poor sample contact and cable bending. As a precursor to an actual clinical implementation, we study the technique in a range of homogeneous liquids with substantially varying dielectric properties. The initial results demonstrate good agreement between the transmission-based probe and commercial, reflection-based probes and pave the way for more substantial clinical implementation.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献