Abstract
The standardization of grayscale display is essentially significant for image signal communication, transmission, and terminal reading. The key step of this standardization is establishing a traceable equipment of grayscale. As a relative value, grayscale is transferred to two different absolute values to satisfy different traceability methods, including optical density for hardcopy image and luminance for softcopy. For luminance, a generation equipment is designed to build the relationship between luminance and grayscale. In this work, novel equipment is established using digital light processing (DLP) by time-frequency modulation, and the corresponding uncertainty is analyzed. The experiment result shows that this digital equipment builds the relationship between grayscale and luminance in the range of 0.16-4000 cd/m2. It enables traceable measurement of grayscale to luminance on this equipment with high accuracy and can provide a standardized reference for the display of grayscale images in the fields of medicine, remote sensing, non-destructive testing, etc.
Funder
National Natural Science Foundation of China
National Key Research and Development Program of China
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献