Abstract
With the rapid development of flexible materials, various high-performance biocompatible flexible sensors have been proposed for specific measurement applications. Among these materials, polydimethylsiloxane (PDMS) is one of the most popular polymers by curing the mixture of pre-polymer (base) and cross-linker (curing agent). In this paper, a force measurement method based on PDMS grating is introduced. The PDMS grating is cast from a commercial master grating, which is precise, low-cost, and easy to follow. The elastic modulus can be controlled by the curing temperature and the mixing ratio. The PDMS grating is tested using a tension testing machine. As the stretching force increases, the grating line-spacing simultaneously increases and the diffraction light spot shifts. By capturing the light spot shift using a camera, the relationship between light spot position and stretching force is established and evaluated. Experimental results show that the linearity (R2) of the proposed method is better than 0.998, adding that the sensitivity is ~0.5–0.7N/mm and the accuracy is up to 0.05N.
Funder
National Natural Science Foundation of China
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
25 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献