Analysis of Friction in Total Knee Prosthesis during a Standard Gait Cycle

Author:

Ranuša Matúš,Wimmer Markus A.ORCID,Fullam SpencerORCID,Vrbka MartinORCID,Křupka Ivan

Abstract

Total knee arthroplasty is on the rise worldwide. Despite its success, revision surgeries are also increasing. According to the American Joint Replacement Registry 2020, 3.3% of revision surgeries are due to wear, and 24.2% are due to mechanical loosening. The combination of shear stresses and wear particles occurring at the bone/implant interface can lead to local osteolysis. Although the shear stresses are partially driven by joint friction, relatively little is known about the evolution of the coefficient of friction (CoF) during a gait cycle in total knee replacement. Here we describe the CoF during a gait cycle and investigate its association with kinematics (slide–roll-ratio), applied load, and relative velocity. The artificial knee was simulated by cobalt–chromium condyle on a flat ultra-high-molecular-weight polyethylene (UHMWPE) tibial plateau, lubricated by either water or proteinaceous solution. We found that the CoF is not a constant but fluctuates between the values close to 0 and 0.15. Cross-correlation suggested that this is primarily an effect of the slide–roll ratio and the contact pressure. There was no difference in the CoF between water and proteinaceous solution. Knowledge about the CoF behavior during a gait cycle will help to increase the accuracy of future computational models of total knee replacement.

Funder

Ministry of Education, Youth and Sports of the Czech Republic

Publisher

MDPI AG

Subject

Surfaces, Coatings and Films,Mechanical Engineering

Reference44 articles.

1. The AJRR 2018 Annual Report: THA/TKA Surgical Insights Summary and Beyond,2018

2. Early Results From the American Joint Replacement Registry: A Comparison With Other National Registries

3. Projections of Primary and Revision Hip and Knee Arthroplasty in the United States from 2005 to 2030

4. National Joint Registry for England, Wales and Northern Ireland;Registry,2017

5. UHMWPE Biomaterials Handbook

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献