Abstract
For long-term and more convenience electrocardiograph (ECG) monitoring, an active- electrode-based ECG monitoring system, which can measure ECG through clothes, is proposed in this paper. The hardware of the system includes active electrodes, signal processing and data transmission modules and the software mainly includes a denoising algorithm based on empirical mode decomposition (EMD). Then the proposed system was verified using the comparison of the ECG signals measured synchronously by active electrodes and Ag/AgCl electrodes. In addition, three flexible materials, including conductive textile, copper foil tape and a flexible printed circuit (FPC) are developed for the sensing layer with active electrodes. To compare the performance of these three materials for the sensing layer, the ECG signals of 10 subjects were measured by different materials in three postures and several indexes for quality evaluation were calculated. Results show that effective and clear ECG waveforms can be measured by all three kinds of materials and the quality of ECG signals measured by FPC is the best by conducting a significant t-test for signal quality indexes of three materials.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
30 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献