Abstract
The main objective of this study is to design titanium alloy femoral stems with cubic porous structures that will be able to reduce stress shielding and promote stem stability. These porous structure designs were introduced into titanium alloy femoral stems as homogeneous and functionally graded porous structures. First, the cubic cellular structures were simulated under compressive loading to measure the yield and modulus of elasticity for various porosity ranges. Based on the selected porosity range, fifteen different arrangements of radial geometrical functionally graded (FG) designs were developed with average porosities of 30, 50, and 70% respectively. Finite element models were developed with physiological loads presenting three different walking speeds (1, 3, and 5 km/h), where the average human body weight was assumed. Stresses at the bone Gruen zones were measured to check the percentage of stress transfer to the bone for each porous stem design and were compared with the bulk stem. Several FG stem designs were shortlisted for further investigation as candidates for hip implants.
Funder
Qatar National Research Fund
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
20 articles.