Evaluation of an Anti-Thrombotic Continuous Lactate and Blood Pressure Monitoring Catheter in an In Vivo Piglet Model undergoing Open-Heart Surgery with Cardiopulmonary Bypass

Author:

Ho Kenneth Kwun YinORCID,Peng Yun-Wen,Ye Minyi,Tchouta Lise,Schneider Bailey,Hayes McKenzie,Toomasian JohnORCID,Cornell Marie,Rojas-Pena Alvaro,Charpie JohnORCID,Chen Hao

Abstract

Blood lactate and blood pressure measurements are important predictors of life-threatening complications after infant open-heart surgeries requiring cardiopulmonary bypass (CPB). We have developed an intravascular nitric oxide (NO)-releasing 5-Fr catheter that contains a lactate sensor for continuous in-blood lactate monitoring and a dedicated lumen for third-party pressure sensor attachment. This device has antimicrobial and antithrombotic properties and can be implanted intravascularly. The importance of this design is its ability to inhibit thrombosis, due to the slow release of NO through the surface of the catheter and around the electrochemical lactate sensors, to allow continuous data acquisition for more than 48 h. An in vivo study was performed using six piglets undergoing open-heart surgery with CPB and cardioplegic arrest, in order to mimic intra-operative conditions for infants undergoing cardiac surgery with CPB. In each study of 3 h, two 5-Fr NO-releasing lactate and blood-pressure monitoring catheters were implanted in the femoral vessels (arteries and veins) and the CPB circuitry to monitor changing lactate levels and blood pressures during and immediately after aortic cross-clamp removal and separation from CBP. Electrical signals continuously acquired through the sensors were processed and displayed on the device’s display and via Bluetooth to a computer in real-time with the use of a two-point in vivo calibration against blood gas results. The study results show that lactate levels measured from those sensors implanted in the CPB circuit during CPB were comparable to those acquired by arterial blood gas measurements, whereas lactate levels measured from sensors implanted in the femoral artery were closely correlated with those acquired intermittently by blood gas prior to CPB initiation, but not during CPB. Blood pressure sensors attached to one lumen of the device displayed accurate blood pressure readings compared to those measured using an FDA approved pressure sensor already on the market. We recommend that the sensor be implanted in the CPB’s circuit to continuously monitor lactate during CPB, and implanted in the femoral arteries or jugular veins to monitor lactate before and after CPB. Blood pressures dramatically drop during CPB due to lower blood flow into the lower body, and we suspect that the femoral arteries are likely collapsing or constricting on the implanted catheter and disrupting the sensor-to-blood contact. This study shows that the device is able to accurately and continuously monitor lactate levels during CPB and potentially prevent post-surgery complications in infants.

Funder

National Institutes of Health

Publisher

MDPI AG

Subject

Physical and Theoretical Chemistry,Analytical Chemistry

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3