Abstract
A sensitive and flexible detection method for organophosphorus pesticides (OPs) detection is a crucial request to avoid their further expanded pollution. Herein, an acetylcholinesterase (AChE) electrochemical sensor, based on the co-modification of ZIF-8 and graphene (GR), was constructed for the detection of OPs. ZIF-8/GR composite can provide a stable and biocompatible environment for the loading of AChE and can accelerate the chemical reaction on the electrode surface. After optimization, the linear detection range of the constructed AChE-CS/GR/ZIF-8/GCE sensor for ICP was 0.5–100 ng/mL (1.73–345.7 nM), and the limit of detection was 0.18 ng/mL (0.62 nM). Moreover, high sensitivity and high specificity of the sensor were also achieved in actual cabbage and tap water samples. Therefore, it has great potential for the application of organophosphorus pesticide residue analysis.
Funder
Independent Scientific Research Project of National University of Defense Technology
Subject
Physical and Theoretical Chemistry,Analytical Chemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献