Author:
Mei Chong Jin,Yusof Nor Azah,Alang Ahmad Shahrul Ainliah
Abstract
This study used a thiolated calix[4]arene derivative modified on gold nanoparticles and a screen-printed carbon electrode (TC4/AuNPs/SPCE) for Pb2+ and Cu2+ determination. The surface of the modified electrode was characterised via Fourier-transform infrared spectroscopy (FTIR), field emission scanning electron microscopy (FESEM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS). Differential pulse voltammetry (DPV) was used for the detection of Pb2+ and Cu2+ under optimum conditions. The limit of detection (LOD) for detecting Pb2+ and Cu2+ was 0.7982 × 10−2 ppm and 1.3358 × 10−2 ppm, respectively. Except for Zn2+ and Hg2+, the presence of competitive ions caused little effect on the current response when detecting Pb2+. However, all competitive ions caused a significant drop in the current response when detecting Cu2+, except Ca2+ and Mg2+, suggesting the sensing platform is more selective toward Pb2+ ions rather than copper (Cu2+) ions. The electrochemical sensor demonstrated good reproducibility and excellent stability with a low relative standard deviation (RSD) value in detecting lead and copper ions. Most importantly, the result obtained in the analysis of Pb2+ and Cu2+ had good recovery in river water, demonstrating the applicability of the developed sensor for real samples.
Funder
Ministry of Higher Education, Malaysia
Subject
Physical and Theoretical Chemistry,Analytical Chemistry
Cited by
25 articles.