Abstract
Computer aided orthopedic surgery suffers from low clinical adoption, despite increased accuracy and patient safety. This can partly be attributed to cumbersome and often radiation intensive registration methods. Emerging RGB-D sensors combined with artificial intelligence data-driven methods have the potential to streamline these procedures. However, developing such methods requires vast amount of data. To this end, a multi-modal approach that enables acquisition of large clinical data, tailored to pedicle screw placement, using RGB-D sensors and a co-calibrated high-end optical tracking system was developed. The resulting dataset comprises RGB-D recordings of pedicle screw placement along with individually tracked ground truth poses and shapes of spine levels L1–L5 from ten cadaveric specimens. Besides a detailed description of our setup, quantitative and qualitative outcome measures are provided. We found a mean target registration error of 1.5 mm. The median deviation between measured and ground truth bone surface was 2.4 mm. In addition, a surgeon rated the overall alignment based on 10% random samples as 5.8 on a scale from 1 to 6. Generation of labeled RGB-D data for orthopedic interventions with satisfactory accuracy is feasible, and its publication shall promote future development of data-driven artificial intelligence methods for fast and reliable intraoperative registration.
Subject
Electrical and Electronic Engineering,Computer Graphics and Computer-Aided Design,Computer Vision and Pattern Recognition,Radiology, Nuclear Medicine and imaging
Cited by
2 articles.