Abstract
High visibility (0.56) neutron-based multi-modal imaging with a Talbot–Lau interferometer at a wavelength of 1.6 Å is reported. A tomography scan of a strongly absorbing quartz geode sample was performed with both the neutron and an X-ray grating interferometer (70 kVp) for a quantitative comparison. Small scattering structures embedded in the absorbing silica matrix were well resolved in neutron dark-field CT slices with a spatial resolution of about 300 μm. Beneficial effects, such as monochromaticity and stronger penetration power of the used neutron radiation, helped to avoid the beam hardening-related artificial dark-field signal which was present in the X-ray data. Both dark-field modalities show mostly the same structures; however, some scattering features appear only in the neutron domain. Potential applications of combined X-ray and neutron multi-modal CT enabling one to probe both the nuclear and the electron density-related structural properties are discussed. strongly absorbing samples are now accessible for the dark-field modality by the use of thermal neutrons.
Funder
Bundesministerium für Bildung und Forschung
Subject
Electrical and Electronic Engineering,Computer Graphics and Computer-Aided Design,Computer Vision and Pattern Recognition,Radiology, Nuclear Medicine and imaging
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Material Discrimination Using X-Ray and Neutron;Journal of Radiation Protection and Research;2023-12-30