Long-Term Clinical Outcomes of Ahmed Valve Implantation in Aniridic Glaucoma

Author:

Bolek Bartłomiej12ORCID,Wylęgała Edward12ORCID,Tarnawska Dorota23ORCID

Affiliation:

1. Chair and Clinical Department of Ophthalmology, School of Medicine in Zabrze, Medical University of Silesia in Katowice, 40-760 Katowice, Poland

2. Clinical Department of Ophthalmology, District Railway Hospital, 40-760 Katowice, Poland

3. Institute of Biomedical Engineering, Faculty of Science and Technology, University of Silesia in Katowice, 75 Pułku Piechoty 1A, 41-500 Chorzów, Poland

Abstract

Background: This study assessed the efficacy and safety of Ahmed valve implantation in patients with aniridic glaucoma for three consecutive years. Methods: Six adult patients (seven eyes) with Ahmed valve (AV) implants for aniridic glaucoma were enrolled in the study. The primary outcome measures were intraocular pressure reduction, glaucoma medication use, success rates, and visual acuity after AV implantation. A 30% reduction in IOP from baseline without the need for re-intervention was considered an effective treatment. The cessation of antiglaucoma medications was defined as complete success. Intraoperative and postoperative complications were included as secondary outcome measures. Measurements were performed preoperatively, at the first week, and 1, 3, 6, 12, 18, 24, 30, and 36 months postoperatively. Results: A total of seven eyes (6 patients) were evaluated 36 months after AV implantation. The mean ± SD values of IOP preoperatively at 1 day, 1 week, and 1, 3, 6, 12, 18, 24, 30, and 36 months postoperatively were 30.4 ± 4.0 mmHg, 14.6 ± 4.6 mmHg, 16.1 ± 4.6 mmHg, 20.7 ± 7.0 mmHg, 14.5 ± 2.7 mmHg, 16.5 ± 5.9 mmHg, 16.2 ± 4.0 mmHg, 16.3 ± 4.3 mmHg, 17.2 ± 10.1 mmHg, 17.6 ± 6.9 mmHg, and 18.2 ± 5.5 mmHg, respectively. At the last follow up, the mean IOP was reduced by 40.2%. The qualified success rate was 85.7%. One patient (one eye) at the last follow-up visit did not require antiglaucoma medications, resulting in a complete success rate of 14.3%. Intra- and postoperative mild or moderate subconjunctival bleeding was observed in all the patients. No other major/minor intraoperative or postoperative complications were noted. Conclusions: In long-term follow up, the AV implantation procedure is well-tolerated and relatively safe for reducing IOP in adult aniridia patients with glaucoma. These results should be validated through studies involving a larger patient cohort.

Funder

Medical University of Silesia in Katowice and University of Silesia in Katowice

Publisher

MDPI AG

Subject

General Biochemistry, Genetics and Molecular Biology,Medicine (miscellaneous)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3