Optimization of Thyroid Volume Determination by Stitched 3D-Ultrasound Data Sets in Patients with Structural Thyroid Disease

Author:

Seifert Philipp1ORCID,Ullrich Sophie-Luise1,Kühnel Christian1ORCID,Gühne Falk1,Drescher Robert1ORCID,Winkens Thomas1ORCID,Freesmeyer Martin1ORCID

Affiliation:

1. Clinic of Nuclear Medicine, Jena University Hospital, Am Klinikum 1, 07747 Jena, Germany

Abstract

Ultrasound (US) is the most important imaging method for the assessment of structural disorders of the thyroid. A precise volume determination is relevant for therapy planning and outcome monitoring. However, the accuracy of 2D-US is limited, especially in cases of organ enlargements and deformations. Software-based “stitching” of separately acquired 3D-US data revealed precise volume determination in thyroid phantoms. The purpose of this study is to investigate the feasibility and accuracy of 3D-US stitching in patients with structural thyroid disease. A total of 31 patients from the clinical routine were involved, receiving conventional 2D-US (conUS), sensor-navigated 3D-US (3DsnUS), mechanically-swept 3D-US (3DmsUS), and I-124-PET/CT as reference standard. Regarding 3DsnUS and 3DmsUS, separately acquired 3D-US images (per thyroid lobe) were merged to one comprehensive data set. Subsequently, anatomical correctness of the stitching process was analysed via secondary image fusion with the I-124-PET images. Volumetric determinations were conducted by the ellipsoid model (EM) on conUS and CT, and manually drawn segmental contouring (MC) on 3DsnUS, 3DmsUS, CT, and I-124-PET/CT. Mean volume of the thyroid glands was 44.1 ± 25.8 mL (I-124-PET-MC = reference). Highly significant correlations (all p < 0.0001) were observed for conUS-EM (r = 0.892), 3DsnUS-MC (r = 0.988), 3DmsUS-MC (r = 0.978), CT-EM (0.956), and CT-MC (0.986), respectively. The mean volume differences (standard deviations, limits of agreement) in comparison with the reference were −10.50 mL (±11.56 mL, −33.62 to 12.24), −3.74 mL (±3.74 mL, −11.39 to 3.78), and 0.62 mL (±4.79 mL, −8.78 to 10.01) for conUS-EM, 3DsnUS-MC, and 3DmsUS-MC, respectively. Stitched 3D-US data sets of the thyroid enable accurate volumetric determination even in enlarged and deformed organs. The main limitation of high time expenditure may be overcome by artificial intelligence approaches.

Publisher

MDPI AG

Subject

General Biochemistry, Genetics and Molecular Biology,Medicine (miscellaneous)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3