Abstract
Among “extra-criteria” antiphospholipid (aPL) antibodies, anti-phosphatidylserine/prothrombin (aPS/PT) antibodies, are considered a part of risk assessment strategies when investigating patients suspected of having antiphospholipid syndrome (APS). aPL detection is currently performed by solid-phase assays to identify anti-cardiolipin (aCL), anti-β2glycoprotein I (aβ2GPI) and aPS/PT antibodies, but new techniques are emerging. Among these, particle-based multi-analyte technology (PMAT), which allows the full automation and simultaneous digital detection of autoantibodies and proteins, including IgG, IgA and IgM isotypes of aCL, aβ2GPI and aPS/PT. The aim of this study was to investigate the agreement of aPS/PT testing between enzyme-linked immunosorbent assay (ELISA) and the PMAT platform. A total of 94 patients were enrolled in the study, including 71 patients with confirmed APS and 23 “aPL carriers”. aPS/PT IgG showed a moderate binomial agreement between ELISA and PMAT (k = 0.57, 95% CI 0.45–0.75), and aPS/PT IgM showed a moderate agreement (k = 0.60, 95% CI 0.45–0.75). Moreover, when considering the continuous agreement, both aPS/PT IgG and IgM showed a statistically significant correlation between ELISA and PMAT (Spearman’s correlation = 0.69, p < 0.001 and 0.72, p < 0.001, respectively). This study demonstrates that PMAT technology is a reliable method for aPS/PT IgG and IgM testing when compared to the available commercial ELISA kit.
Subject
General Biochemistry, Genetics and Molecular Biology,Medicine (miscellaneous)
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献