Abstract
Electromagnetic bi-directional micropumps (EMBM) are indispensable for the development of portable devices which enable fluid transportation in forward and reverse directions. However, the high disposal cost of rare-earth magnets attached to the fluidic part and the low pump density due to multiple motors limit their practical application in disposable multi-channel microfluidic applications such as droplet-based oscillatory-flow (DBOF) rapid PCR amplification. Therefore, this paper presented a low-cost, disposable, high-pump-density EMBM. To reduce the disposal cost, we separated the magnets from the disposable fluidic part and used cylindrical holes to store and guide the magnet, which resulted in the ability to reuse all the magnets. To increase the pump density, we used the combination of one motor and one multi-pole ring magnet to drive several channels simultaneously. A proof-of-concept prototype with a pump density of 0.28 cm−2 was fabricated and experimentally evaluated. The fabricated micropump exhibited a maximum flow rate of 0.86 mL/min and a maximum backpressure of 0.5 kPa at a resonant frequency around 50 Hz. The developed multi-channel micropump with reusable magnets is highly beneficial to the development of low-cost and high-throughput rapid PCR amplification microchips and therefore can have a significant impact on timely infectious disease recognition and intervention.