Comparison of Tandem Mass Spectrometry and the Fluorometric Method—Parallel Phenylalanine Measurement on a Large Fresh Sample Series and Implications for Newborn Screening for Phenylketonuria

Author:

Perko Dasa1ORCID,Groselj Urh2ORCID,Cuk Vanja1,Iztok Remec Ziga1,Zerjav Tansek Mojca23,Drole Torkar Ana2,Krhin Blaz4,Bicek Ajda4,Oblak Adrijana4ORCID,Battelino Tadej23,Repic Lampret Barbka1

Affiliation:

1. Clinical Institute for Special Laboratory Diagnostics, University Children’s Hospital, UMC Ljubljana, 1000 Ljubljana, Slovenia

2. Department of Endocrinology, Diabetes and Metabolic Diseases, University Children’s Hospital, UMC Ljubljana, 1000 Ljubljana, Slovenia

3. Faculty of Medicine, University of Ljubljana, 1000 Ljubljana, Slovenia

4. Department of Nuclear Medicine, UMC Ljubljana, 1000 Ljubljana, Slovenia

Abstract

Phenylketonuria (PKU) was the first disease to be identified by the newborn screening (NBS) program. Currently, there are various methods for determining phenylalanine (Phe) values, with tandem mass spectrometry (MS/MS) being the most widely used method worldwide. We aimed to compare the MS/MS method with the fluorometric method (FM) for measuring Phe in the dried blood spot (DBS) and the efficacy of both methods in the NBS program. The FM was performed using a neonatal phenylalanine kit and a VICTOR2TM D fluorometer. The MS/MS method was performed using a NeoBaseTM 2 kit and a Waters Xevo TQD mass spectrometer. The Phe values measured with the MS/MS method were compared to those determined by the FM. The cut-off value for the NBS program was set at 120 µmol/L for FM and 85 µmol/L for MS/MS. We analyzed 54,934 DBS. The measured Phe values varied from 12 to 664 µmol/L, with a median of 46 µmol/L for the MS/MS method and from 10 to 710 µmol/L, with a median of 70 µmol/L for the FM. The Bland–Altman analysis indicated a bias of −38.9% (−23.61 µmol/L) with an SD of 21.3% (13.89 µmol/L) when comparing the MS/MS method to the FM. The Phe value exceeded the cut-off in 187 samples measured with FM and 112 samples measured with MS/MS. The FM had 181 false positives, while the MS/MS method had 106 false positives. Our study showed that the MS/MS method gives lower results compared to the FM. Despite that, none of the true positives would be missed, and the number of false-positive results would be significantly lower compared to the FM.

Funder

Slovenian Research Agency

Publisher

MDPI AG

Subject

Inorganic Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Computer Science Applications,Spectroscopy,Molecular Biology,General Medicine,Catalysis

Reference38 articles.

1. Newborn screening for inherited metabolic disorders; news and views;Pourfarzam;J. Res. Med. Sci.,2013

2. Expanded newborn screening program in Slovenia using tandem mass spectrometry and confirmatory next generation sequencing genetic testing;Remec;Zdr. Varst.,2020

3. The Phenylalanine Hydroxylating System;Kaufman;Adv. Enzymol. Relat. Areas Mol. Biol.,2006

4. Genetske in klinične značilnosti bolnikov s fenilketonurijo v Sloveniji;Battelino;Zdr. Vestn.,2013

5. Screening tests: A review with examples;Maxim;Inhal. Toxicol.,2014

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3