Non-Contact Real-Time Measurement of Refractive Index of Flowing Liquid Based on Optical Coherence Tomography

Author:

Fu Hongbo1,Gao Weijian1,Lin Zixin1,Zeng Zhemin1,Shi Wen1,Zhang Jian1ORCID

Affiliation:

1. School of Biomedical Engineering, Guangzhou Medical University, Guangzhou 511436, China

Abstract

The refractive index (RI) is one of the liquid’s most important physical parameters of liquid. The change of RI usually indicates the change of liquid properties, and this relation leads to broad application. This paper proposes a method based on optical coherence tomography (OCT) to measure the RI of liquid in real time. A series of targeted image-processing techniques play a vital role in this method. Benefiting from the advantages of OCT, this method does not require the operators to come into direct contact with liquids. Moreover, the principles and techniques adopted in this method imply that the measuring range of RI of unknown liquids is unlimited. The effectiveness of this method is validated by the experiment measuring the RI of sucrose liquids with gradient concentration. Another experiment thoroughly verifies this method’s feasibility in predicting sucrose concentration based on real-time measured RI. This simple and easy proposed method indicates an immense potential application prospect in the industry and scientific research.

Funder

National Key R&D Program of China

Publisher

MDPI AG

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Atomic and Molecular Physics, and Optics

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3