Abstract
Structured illumination microscopy (SIM) is widely applied due to its high temporal and spatial resolution imaging ability. sCMOS cameras are often used in SIM due to their superior sensitivity, resolution, field of view, and frame rates. However, the unique single-pixel-dependent readout noise of sCMOS cameras may lead to SIM reconstruction artefacts and affect the accuracy of subsequent statistical analysis. We first established a nonuniform sCMOS noise model to address this issue, which incorporates the single-pixel-dependent offset, gain, and variance based on the SIM imaging process. The simulation indicates that the sCMOS pixel-dependent readout noise causes artefacts in the reconstructed SIM superresolution (SR) image. Thus, we propose a novel sCMOS noise-corrected SIM reconstruction algorithm derived from the imaging model, which can effectively suppress the sCMOS noise-related reconstruction artefacts and improve the signal-to-noise ratio (SNR).
Funder
National Natural Science Foundation of China
National Key Research and Development Programme of China
Beijing Natural Science Foundation
Natural Science Foundation of Chongqing
Subject
Radiology, Nuclear Medicine and imaging,Instrumentation,Atomic and Molecular Physics, and Optics
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献