Abstract
The micropillar array electrode (µAE) has been widely applied in microchip-based electrochemical detection systems due to a large current response. However, it was found that amplifying the current through further adjusting geometrical parameters is generally hindered by the shielding effect. To solve this problem, a bio-inspired micropillar array electrode (bµAE) based on the microfluidic device has been proposed in this study. The inspiration is drawn from the structure of leatherback sea turtles’ mouths. By deforming a μAE to rearrange the micropillars on bilateral sides of the microchannel, the contact area between micropillars and analytes increases, and thus the current is substantially improved. A numerical simulation was then used to characterize the electrochemical performance of bµAEs. The effects of geometrical and hydrodynamic parameters on the current of bµAEs were investigated. Moreover, a prototypical microchip integrated with bµAE was fabricated for detailed electrochemical measurement. The chronoamperometry measurements were conducted to verify the theoretical performance of bµAEs, and the results suggest that the experimental data are in good agreement with those of the simulation model. This work presents a novel bµAE with great potential for highly sensitive electrochemical detection and provides a new perspective on the efficient configuration of the µAE.
Funder
the Shenzhen Science and Technology Innovation Committee
Subject
Clinical Biochemistry,General Medicine,Analytical Chemistry,Biotechnology,Instrumentation,Biomedical Engineering,Engineering (miscellaneous)
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献