Abstract
The elliptical paraboloid array plays an important role in precision measurement, astronomical telescopes, and communication systems. The calibration of the vertex distance of elliptical paraboloids is of great significance to precise 2D displacement measurement. However, there are some difficulties in determining the vertex position with contact measurement. In this study, an elliptical paraboloid array and an optical slope sensor for displacement measurement were designed and analyzed. Meanwhile, considering the geometrical relationship and relative angle between elliptical paraboloids, a non-contact self-calibration method for the vertex distance of the elliptical paraboloid array was proposed. The proposed self-calibration method was verified by a series of experiments with a high repeatability, within 3 μ m in the X direction and within 1 μ m in the Y direction. Through calibration, the displacement measurement system error was reduced from 100 μ m to 3 μ m . The self-calibration method of the elliptical paraboloid array has great potential in the displacement measurement field, with a simple principle and high precision.
Funder
Natural Science Foundation of Anhui Province
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献