Author:
Mark Erwin P.,Lewis Michael A. O.,Graziani Filippo,Atlas Boris,Utsch Joern
Abstract
The COVID-19 pandemic has drawn attention to microbial transmission risk via aerosols in dental practice. Demonstration electric toothbrushes are used intra-orally for education. The aim of this investigation was to measure the size of droplets emitted by the brush head of two demonstration oscillating-rotating electric toothbrushes. Measurement of droplet production and size was recorded in vitro using three methods: (1) Malvern Spraytec (LASER particle size measurement device with detectable particle size of 0.1–2500 µm) and brushes mounted on a 3D-printed, two-shell form-fit fixture with a supply of tap water; (2) a DustTrak aerosol measurement device and toothpaste slurry, with brushing simulated in the oral cavity of a phantom head; (3) high-speed visualization in a simulated-use situation in the oral cavity of a phantom head, with individual evaluation of tap water, water with detergent, 70% ethanol, glycerin and toothpaste slurry. Both brushes showed the size of emitted droplets was consistently between 200 and 1200 µm, categorized as splatter (dental aerosols are <50 µm diameter). No significant incremental aerosol-sized matter was detected during toothbrush operation. The high-speed video visualization confirmed only splatter-sized droplets during operation. These findings indicate that oscillating-rotating toothbrushes do not produce aerosol-sized particles during simulated use.
Funder
Procter & Gamble Service GmbH
Subject
Health, Toxicology and Mutagenesis,Public Health, Environmental and Occupational Health