Author:
Martini Francesco,Giovine Ennio,Chiarello Fabio,Carelli Pasquale
Abstract
We describe a THz spectrometer operating between 1.2 and 10.5 THz, consisting of band pass filters made by metasurfaces. The source is made of a 10 W small black body. The detector is a high sensitivity room temperature pyroelectric sensor. Various techniques that are used to prepare samples are described. The spectra obtained are compared with those measured with a Fourier Transformer Infrared Spectrometer on the same samples. Our instrument, which uses commercial technologies available at the present time, can constitute an economical alternative to very expensive spectrometers. It has already been successfully used, obtaining precise spectroscopic measurements on many inorganic powders.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献