Abstract
Electrical impedance tomography (EIT) is a non-invasive detection technology that uses the electrical response value at the boundary of an observation field to image the conductivity changes in an area. When EIT is applied to the thoracic cavity of the human body, the conductivity change caused by the heartbeat will be concentrated in a sub-region of the thoracic cavity, that is, the heart region. In order to improve the spatial resolution of the target region, two sensor optimization methods based on conformal mapping theory were proposed in this study. The effectiveness of the proposed method was verified by simulation and phantom experiment. The qualitative analysis and quantitative index evaluation of the reconstructed image showed that the optimized model could achieve higher imaging accuracy of the heart region compared with the standard sensor. The reconstruction results could effectively reflect the periodic diastolic and systolic movements of the heart and had a better ability to recognize the position of the heart in the thoracic cavity.
Funder
National Natural Science Foundation of China
Natural Science Foundation of Tianjin
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry