Abstract
We evaluated the benefits of the MotionFree algorithm through phantom and patient studies. The various sizes of phantom and vacuum vials were linked to RPM moving with or without MotionFree application. A total of 600 patients were divided into six groups by breathing protocols and CT scanning time. Breathing protocols were applied as follows: (a) patients who underwent scanning without any breathing instructions; (b) patients who were instructed to hold their breath after expiration during CT scan; and (c) patients who were instructed to breathe naturally. The length of PET/CT misregistration was measured and we defined the misregistration when it exceeded 10 mm. In the phantom tests, the images produced by the MotionFree algorithm were observed to have excellent agreement with static images. There were significant differences in PET/CT misregistration according to CT scanning time and each breathing protocol. When applying the type (c) protocol, decreasing the CT scanning time significantly reduced the frequency and length of misregistrations (p < 0.05). The MotionFree application is able to correct respiratory motion artifacts and to accurately quantify lesions. The shorter time of CT scan can reduce the frequency, and the natural breathing protocol also decreases the lengths of misregistrations.
Funder
VHS Medical Center Research Grant
Subject
General Earth and Planetary Sciences,General Environmental Science
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献