Abstract
Bag-valve-mask is a device that manually provides positive oxygen pressure. The grip technique of the character E and C shape is recommended to carry out this effectively. However, when applying this method, the hand in which the direction of the EC technique should be performed and the degree of adhesion pressure while performing the technique are unknown. This study aims to identify the factors influencing tidal volume and to determine the ideal sealing method between mask and face in the one-handed EC technique to optimize the Vt. A simulation study was conducted using a mechanical lung model in a scenario that resembled respiratory arrest. Multiple regression analyses identified high peak pressure, high left spot adhesion strength of the mask, and low right spot and bottom spot adhesion strength of the mask as significant factors. To provide an optimal Vt, it may be necessary to apply more strength to the left area of the mask when forming the “C” shape with the thumb and index finger of left hand using the one-handed EC technique.
Subject
Health Information Management,Health Informatics,Health Policy,Leadership and Management
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献