Author:
Koltermann Jan Jens,Gerber Martin
Abstract
In this scientific study, the question of the influence of the quantization error on the CoP measurement is be clarified. For this purpose, the quantization error is investigated in two scenarios, first with the technical/physical reproduction of the CoP, and then with test persons. From the results, a model is derived with which a technical and economic optimum between resolution and error can be generated for an individual case. The study was carried out with 170 healthy volunteers, aged 20–30 years. The test persons stood in a bipedal position for 15 s on a Kislter force plate (type 9260AA). In the investigation, it was shown that, for the measurement of center of pressure (CoP), signals to mostly 16-bit analog/digital converters are suitable but not, per se, the most economical variant. With the introduction of a quality criterion, a reasonable design for the planned test case can be made.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献