How Does the Micro-Groove Profile Influence the Mechanics of Taper Junction in Hip Implants? A Finite Element Study

Author:

Kalwar Akash1,Feyzi Mohsen1,Hashemi Reza1ORCID

Affiliation:

1. College of Science and Engineering, Flinders University, Tonsley, SA 5042, Australia

Abstract

This study aims to investigate the effect of ridged (micro-grooved) surface finish over the trunnion surface on the mechanics (stress, strain, and deformation) of the head–neck taper interface in hip implants. Using finite element modelling, the study focused on the geometric parameters of such micro-grooves to study how they would mechanically affect stress and deformation fields after the assembly procedure. As such, five different 2D models with varying micro-groove height and spacing were produced and assembled under an impaction assembly force of 4 kN in a 32 mm CoCrMo head engaged with a 12/14 Ti-6Al-4V neck. The results showed that lower von Mises stresses could be induced by either an increase or decrease in spacing against the base model (Model 1), which probably signifies that the relationship between the ridge spacing and stress may depend on the level of spacing. It was concluded that the geometrical parameters of the ridges (and their non-linear interactions) impact not only the stress and strain fields but also the assembly loading time at which the maximal stress and plastic deformation occur.

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3