Author:
Kim Mi,Lee Sungwoo,Yoon Inug,Kook Geon,Jung Yeon,Bawazir Sarah,Stefanini Cesare,Lee Hyunjoo
Abstract
Chronic monitoring of bladder activity and urine volume is essential for patients suffering from urinary dysfunctions. However, due to the anatomy and dynamics of the bladder, chronic and precise monitoring of bladder activity remains a challenge. Here, we propose a new sensing mechanism that measures the bladder volume using a resistive ladder network with contact switches. Instead of measuring the impedance between the electrode continuously, the proposed sensor provides a digitized output (‘on’ or ‘off’) when the bladder volume reaches a certain threshold value. We present simple proof-of-concept sensors which compare the discrete-mode operation to the continuous-mode operation. In addition, by using multiple pairs of this contact-mode switch in a resistor ladder structure, we demonstrate monitoring of the bladder volume in four discrete steps using an idealized balloon and an ex vivo pig’s bladder. We implemented the resistive ladder network using a conductive polypyrrole/agarose hydrogel composite which exhibits a Young’s modulus comparable to that of the bladder wall. Compared to the continuous-mode operation, the proposed sensing mechanism is less susceptible to drift due to material degradation and environmental factors.
Funder
National Research Foundation of Korea
Ministry of Science and ICT
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献