Abstract
A novel and simple optical biosensor to detect triglycerides (TGs) has been successfully constructed by using pectin hydrogel membrane as the indicator pH and chromoionophore ETH 5294 (CI), with lipase as the catalyst. The enzymatic working system against TGs releasing H+ ions will affect the color absorbance of CI. The characterization results show that a TG biosensor has the optimum condition and sensitivity at the phosphate buffer concentration of 50 mM, pH 7, and enzyme loading of 60 μg. The biosensor works at the tripalmitin (TP) concentration range of 100–400 mg/dL. With the sensitivity of 0.001 (∆A/(mg/dL)), the biosensor response reaches stability after five minutes, and the limit of detection (LOD) of the TG optical biosensor is 15 mg/dL. Relative standard deviation (RSD) in a reproducibility test was 2.5%, with a 15-day lifespan.
Funder
Kementerian Riset Teknologi Dan Pendidikan Tinggi Republik Indonesia
Subject
Clinical Biochemistry,General Medicine
Cited by
48 articles.
订阅此论文施引文献