Fibre Bragg Grating Based Interface Pressure Sensor for Compression Therapy

Author:

Bradbury James A.ORCID,Zhang QimeiORCID,Hernandez Ledezma Francisco U.ORCID,Correia RicardoORCID,Korposh SerhiyORCID,Hayes-Gill Barrie R.ORCID,Tamoué Ferdinand,Parnham Alison,McMaster Simon A.,Morgan Stephen P.ORCID

Abstract

Compression therapy is widely used as the gold standard for management of chronic venous insufficiency and venous leg ulcers, and the amount of pressure applied during the compression therapy is crucial in supporting healing. A fibre optic pressure sensor using Fibre Bragg Gratings (FBGs) is developed in this paper to measure sub-bandage pressure whilst removing cross-sensitivity due to strain in the fibre and temperature. The interface pressure is measured by an FBG encapsulated in a polymer and housed in a textile to minimise discomfort for the patient. The repeatability of a manual fabrication process is investigated by fabricating and calibrating ten sensors. A customized calibration setup consisting of a programmable translation stage and a weighing scale gives sensitivities in the range 0.4–1.5 pm/mmHg (2.6–11.3 pm/kPa). An alternative calibration method using a rigid plastic cylinder and a blood pressure cuff is also demonstrated. Investigations are performed with the sensor under a compression bandage on a phantom leg to test the response of the sensor to changing pressures in static situations. Measurements are taken on a human subject to demonstrate changes in interface pressure under a compression bandage during motion to mimic a clinical application. These results are compared to the current gold standard medical sensor using a Bland–Altman analysis, with a median bias ranging from −4.6 to −20.4 mmHg, upper limit of agreement (LOA) from −13.5 to 2.7 mmHg and lower LOA from −32.4 to −7.7 mmHg. The sensor has the potential to be used as a training tool for nurses and can be left in situ to monitor bandage pressure during compression therapy.

Funder

Engineering and Physical Sciences Research Council

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Reference37 articles.

1. Providing optimal sub-bandage pressure in compression therapy;Barrett;J. Community Nurs.,2015

2. Effective compression therapy how to guide;Vowden;Wound Essent.,2012

3. Approaches to the application and removal of compression therapy: A literature review

4. Venous leg ulcer management in clinical practice in the UK: costs and outcomes

5. Cost of managing patients with venous leg ulcers

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献