Abstract
There is increasing evidence that pelvic mobility is a critical factor to consider in implant alignment during total hip arthroplasty (THA). Here, we test the feasibility of using an inertial sensor fitted across the sacrum to measure change in pelvic tilt, and hence screen for patients with high pelvic mobility. Patients (n = 32, mean age: 57.4 years) due to receive THA surgery participated in the study. Measures of pelvic tilt were captured simultaneously using the device and radiograph in three functional positions: Standing, flexed-seated, and step-up. We found a strong correlation between the device and radiograph measures for the change in pelvic tilt measure from standing to flexed-seated position (R2 = 0.911); 75% of absolute errors were under 5 degrees. We demonstrated that the device can be used as a screening tool to rapidly identify patients who would benefit from more detailed surgical planning of implant positioning to reduce future risks of impingement and dislocation.