Abstract
This study aimed to compare the accuracy of three types of apex locators versus digital radiography for working length (WL) determination. This experimental study was conducted on 58 extracted maxillary premolars. The teeth were decoronated, the access cavity was prepared, and WL was determined using a #15 K-file to serve as reference. The WL was then measured by Woodpex V, Woodpex III, and Root ZX apex locators in the presence of 0.9% saline, and also on a photostimulable phosphor plate (PSP) digital radiograph taken by the parallel technique. The values were compared with the actual WL using the paired t-test (alpha = 0.05). Digital radiography, Root ZX, Woodpex V, and Woodpex III determined the WL within ±0.5 mm from the actual value in 84.48%, 100%, 89.66%, and 87.93% of the cases, respectively. Woodpex V (p = 0.039), Woodpex III (p = 0.001), and Root ZX (p = 0.001) significantly over-estimated the WL. The WL measured on digital radiographs was not significantly different from the actual WL (p = 0.213). The position of the apical foramen (central/lateral) had no significant effect on the accuracy of WL determination by different techniques (p >0.05). Within the limitations of this in vitro study, all the tested modalities showed acceptable accuracy for WL determination in maxillary premolars.
Reference34 articles.
1. A Comparative Evaluation of Four Different Techniques for Determining the Accuracy of Root Canal Working Length: An In Vitro Study;Yadav;Dent. Hypotheses,2020
2. De Santis, R., Iaculli, F., Lodato, V., Gallicchio, V., Simeone, M., Spagnuolo, G., and Rengo, C. (2022). The Efficacy of Selected Sodium Hypochlorite Heating Methods for Increasing and Maintaining Its Intracanal Temperature—An Ex Vivo Study. Appl. Sci., 12.
3. Chromographic Analysis and Cytotoxic Effects of Chlorhexidine and Sodium Hypochlorite Reaction Mixtures;Nocca;J. Endod.,2017
4. Comparison of Digital Radiography, Conventional Film and Self-Developing Film for Working Length Determination;Sameye;Iran. Endod. J.,2018
5. Preoperative Estimation of Endodontic Working Length with Cone-Beam Computed Tomography and Standardized Paralleling Technique in comparison to Its Real Length;Faraj;BioMed Res. Int.,2020
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献