Abstract
We developed an implantable electrocardiogram (ECG) monitoring system and demonstrated its performance through an in vivo test. In the system, the implantable device senses not only the ECG signal of the animal but also the voltage level of the secondary cell and temperature inside the implantable device, and users can check the transmitted information through a PC program or a mobile application. The adoption of wireless charging technology eliminates the use of a lead wire and repetitive surgery to replace the implantable device. The proposed wireless charging technology demonstrated experimentally a wireless power transfer efficiency of approximately 30%. To minimize the size of the implantable device, the antenna and coil were integrated into a size of 34 mm × 14 mm. Communication between the implantable device and the basestation can reach up to 2.4 m when the implantable device is inserted into a porcine skin sample.
Funder
National Research Foundation of Korea
Electronics and Telecommunications Research Institute
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献