Abstract
Magnetic induction tomography (MIT) is largely focused on applications in biomedical and industrial process engineering. MIT has a great potential for imaging metallic samples; however, there are fewer developments directed toward the testing and monitoring of metal components. Eddy-current non-destructive testing is well established, showing that corrosion, fatigue and mechanical loading are detectable in metals. Applying the same principles to MIT would provide a useful imaging tool for determining the condition of metal components. A compact MIT instrument is described, including the design aspects and system performance characterisation, assessing dynamic range and signal quality. The image rendering ability is assessed using both external and internal object inclusions. A multi-frequency MIT system has similar capabilities as transient based pulsed eddy current instruments. The forward model for frequency swap multi-frequency is solved, using a computationally efficient numerical modelling with the edge-based finite elements method. The image reconstruction for spectral imaging is done by adaptation of a spectrally correlative base algorithm, providing whole spectrum data for the conductivity or permeability.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry