Abstract
Detecting dengue virus (DENV) infection in patients as early as possible makes the disease management convenient. Conventionally, DENV infection is diagnosed by ELISA-based methods, but sensitivity and specificity are major concerns. Reverse-transcription-PCR (RT-PCR)-based detection confirms the presence of DENV RNA; however, it is expensive, time-consuming, and skilled personnel are required. A fluorescence-based detection system that detects DENV RNA in patient’s serum directly, without any nucleic acid amplification step, has been developed. The method uses target-specific complementary sequence in the molecular beacon, which would specifically bind to the DENV RNA. The molecular beacons are approximately 40 bases long hairpin structures, with a fluorophore-quencher system attached at the terminal ends of the stem. These probes are biotinylated in the stem region, so that they can be immobilized on the streptavidin-tagged magnetic beads. These magnetic beads, coupled with biotinylated molecular beacons, are used for the detection of the target RNA in the serum by incubating the mixture. After incubation, beads are separated and re-suspended in a buffer. The measurement of fluorescence is taken in fluorometer after 15 min incubation at 50 °C. The whole work is carried out in a single tube. This rapid method can precisely detect dengue RNA within two hours, confirming ongoing DENV replication in the patient.