Author:
Venkataraman Abinaya Priya,Sirak Delila,Brautaset Rune,Dominguez-Vicent Alberto
Abstract
Objective: To evaluate the performance of two subjective refraction measurement algorithms by comparing the refraction values, visual acuity, and the time taken by the algorithms with the standard subjective refraction (SSR). Methods: The SSR and two semi-automated algorithm-based subjective refraction (SR1 and SR2) in-built in the Vision-R 800 phoropter were performed in 68 subjects. In SR1 and SR2, the subject’s responses were recorded in the algorithm which continuously modified the spherical and cylindrical component accordingly. The main difference between SR1 and SR2 is the use of an initial fogging step in SR1. Results: The average difference and agreement limits intervals in the spherical equivalent between each refraction method were smaller than 0.25 D, and 2.00 D, respectively. For the cylindrical components, the average difference was almost zero and the agreement limits interval was less than 0.50 D. The visual acuities were not significantly different among the methods. The times taken for SR1 and SR2 were significantly shorter, and SR2 was on average was three times faster than SSR. Conclusions: The refraction values and the visual acuity obtained with the standard subjective refraction and algorithm-based methods were similar on average. The algorithm-based methods were significantly faster than the standard method.