Abstract
We report the design simulation of the Raman spectrometer using Zemax optical system design software. The design is based on the Czerny–Turner configuration, which includes an optical system consisting of an entrance slit, two concave mirrors, reflecting type diffraction grating and an image detector. The system’s modeling approach is suggested by introducing the corresponding relationship between detector pixels and wavelength, linear CCD receiving surface length and image surface dimension. The simulations were carried out using the POP (physical optics propagation) algorithm. Spot diagram, relative illumination, irradiance plot, modulation transfer function (MTF), geometric and encircled energy were simulated for designing the Raman spectrometer. The simulation results of the Raman spectrometer using a 527 nm wavelength laser as an excitation light source are presented. The present optical system was designed in sequential mode and a Raman spectrum was observed from 530 nm to 630 nm. The analysis shows that the system’s image efficiency was quite good, predicting that it could build an efficient and cost-effective Raman spectrometer for optical diagnostics.