Immunoassay System Based on the Technology of Time-Resolved Fluorescence Resonance Energy Transfer
Author:
Xu Zhengping1, Zhou Hong1, Li Li1ORCID, Chen Zhang1, Zhang Xin1, Feng Yongtong1, Wang Jianping1, Li Yuan2, Wu Yanfan1
Affiliation:
1. Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou 215163, China 2. Chongqing Guoke Medical Innovation Technology Development Co., Ltd., Chongqing 401122, China
Abstract
To enhance the specificity and sensitivity, cut the cost, and realize joint detection of multiple indicators, an immunoassay system based on the technology of time-resolved fluorescence resonance energy transfer (TR-FRET) was studied. Due to the FRET of the reagent, the donor probe and acceptor probe emitted specific fluorescence to enhance specificity. Long-lifetime specific fluorescence from the acceptor probe was combined with time-resolved technology to enhance sensitivity. A xenon flash lamp and a photomultiplier tube (PMT) were selected as the light source and detector, respectively. A filter-switching mechanism was placed in the light path, so the fluorescence signal from the donor and acceptor was measured alternately. The instrument’s design is given, and some specificI parts are described in detail. Key technical specifications of the instrument and procalcitonin (PCT), C-reactive protein (CRP), and interleukin-6(IL-6) were tested, and the test results were presented subsequently. The CV value of the self-designed counting module is better than 0.01%, and the instrument noises for 620 nm and 665 nm are 41.44 and 10.59, respectively. When set at 37 °C, the temperature bias (B) is 0.06 °C, and the temperature fluctuation is 0.10 °C. The CV and bias are between ±3% and 5%, respectively, when pipetting volumes are between 10 μL and 100 μL. Within the concentration range of 0.01 nM to 10 nM, the luminescence values exhibit linear regression correlation coefficients greater than 0.999. For PCT detection, when the concentration ranges from 0.02 ng/mL to 50 ng/mL, the correlation coefficient of linear fitting exceeds 0.999, and the limit of quantification is 0.096 ng/mL. For CRP and IL-6, the detection concentration ranges from 0 ng/mL to 500 ng/mL and 0 ng/mL to 20 ng/mL, respectively, with limits of quantification of 2.70 ng/mL and 2.82 ng/mL, respectively. The experimental results confirm the feasibility of the technical and instrumental solutions.
Funder
the Management Committee of Chongqing Liangjiang New Area
Reference38 articles.
1. Rohr, U.-P., Binder, C., Dieterle, T., Giusti, F., Messina, C.G.M., Toerien, E., Moch, H., and Schäfer, H.H. (2016). The Value of In Vitro Diagnostic Testing in Medical Practice: A Status Report. PLoS ONE, 11. 2. Paper-based fluorogenic devices for in vitro diagnostics;Wu;Biosens. Bioelectron.,2018 3. Wang, L., Xu, W., Wang, B., Si, X., and Li, S. (2023). Methods and Advances in the Design, Testing and Development of In Vitro Diagnostic Instruments. Processes, 11. 4. The in vitro diagnostics industry in China;Song;View,2020 5. Chemiluminescent immunoassay technology: What does it change in autoantibody detection?;Cinquanta;Autoimmun. Highlights,2017
|
|