Exploring Stresses in Mandibular Jawbone during Implant Insertion: A Three-Dimensional Explicit Dynamic Analysis

Author:

K N Chethan1ORCID,Eram Afiya2ORCID,Shetty Nisha3,Shetty Divya D.1ORCID,Futane Mohan4,Keni Laxmikant G.1ORCID

Affiliation:

1. Department of Aeronautical and Automobile Engineering, Manipal Institute of Technology, Manipal, Manipal Academy of Higher Education, Manipal 576104, Karnataka, India

2. Department of Conservative Dentistry and Endodontics, Manipal College of Dental Sciences, Manipal, Manipal Academy of Higher Education, Manipal 576104, Karnataka, India

3. Department of Oral and Maxillofacial Pathology and Oral Microbiology, Manipal College of Dental Sciences, Manipal, Manipal Academy of Higher Education, Manipal 576104, Karnataka, India

4. Department of Mechanical Engineering, Hirasugar Institute of Technology, Nidasoshi, Belagavi 591236, Karnataka, India

Abstract

In dental implant insertion, an artificial foundation is prepared for the prosthetic device, which involves the surgical positioning of the implant in the jaw bone. The success of dental implants relies on the osseointegration process. The biomechanical factors, such as stress and strain, developed during the insertion affect the jawbone and its surroundings. In this current study, the stresses during the implant insertion in the mandibular jawbone bone are analyzed using three-dimensional explicit dynamic analysis, and the Cowper–Symonds model is implemented with the damage model. The implant’s design has a substantial impact on stress distribution within the cancellous bone during the insertion procedure. The stress variation takes place as the implant moves into the pre-drilled hole. This is because of the contact between the bone and the fixture on the implant. The upper edge of the predrilled site shows that the stresses are more at the crestal region of the implant due to surface area. There is a gradual increase in the stress level as the implant reaches the lower edge from the top edge. This is because of the concept of mechanical interlocking. Clinicians can use this information to anticipate and address potential stress-related challenges during implant placement.

Funder

Manipal Academy of Higher Education