A 10 V Transfer Standard Based on Low-Noise Solid-State Zener Voltage Reference ADR1000
-
Published:2024-03-05
Issue:1
Volume:4
Page:98-116
-
ISSN:2673-8244
-
Container-title:Metrology
-
language:en
-
Short-container-title:Metrology
Author:
Bülau André1ORCID, Walter Daniela12ORCID, Zimmermann André12ORCID
Affiliation:
1. Hahn-Schickard, Allmandring 9b, 70569 Stuttgart, Germany 2. Institute for Micro Integration (IFM), University of Stuttgart, Allmandring 9b, 70569 Stuttgart, Germany
Abstract
Voltage standards are widely used to transfer volts from Josephson voltage standards (JVSs) at national metrology institutes (NMIs) into calibration labs to maintain the volts and to transfer them to test equipment at production lines. Therefore, commercial voltage standards based on Zener diodes are used. Analog Devices Inc. (San Jose, CA, USA), namely, Eric Modica, introduced the ADR1000KHZ, a successor to the legendary LTZ1000, at the Metrology Meeting 2021. The first production samples were already available prior to this event. In this article, this new temperature-stabilized Zener diode is compared to several others as per datasheet specifications. Motivated by the superior parameters, a 10 V transfer standard prototype for laboratory use with commercial off-the-shelf components such as resistor networks and chopper amplifiers was built. How much effort it takes to reach the given parameters was investigated. This paper describes how the reference was set up to operate it at its zero-temperature coefficient (z.t.c.) temperature and to lower the requirements for the oven stability. Furthermore, it is shown how the overall temperature coefficient (t.c.) of the circuit was reduced. For the buffered Zener voltage, a t.c. of almost zero, and with amplification to 10 V, a t.c. of <0.01 µV/V/K was achieved in a temperature span of 15 to 31 °C. For the buffered Zener voltage, a noise of ~584 nVp-p and for the 10 V output, ~805 nVp-p were obtained. Finally, 850 days of drift data were taken by comparing the transfer standard prototype to two Fluke 7000 voltage standards according to the method described in NBS Technical Note 430. The drift specification was, however, not met.
Reference46 articles.
1. The ampere and electrical standards;Elmquist;J. Res. Natl. Inst. Stand. Technol.,2001 2. Hamer, W.J. (1965). Standard Cells: Their Construction, Maintenance, and Characteristics. 3. The units for mass, voltage, resistance, and electrical current in the SI;Schlamminger;IEEE Instrum. Meas. Mag.,2019 4. Rüfenacht, A., Fox, A.E., Butler, G.E., Burroughs, C.J., Dresselhaus, P.D., Schwall, R.E., Cular, S., and Benz, S.P. (2020, January 24–28). Compact DC Josephson Voltage Standard. Proceedings of the 2020 Conference on Precision Electromagnetic Measurements (CPEM), Denver, CO, USA. 5. Maruyama, M., Urano, C., Kaneko, N.-H., Sannomaru, E., Yonezawa, T., Kanai, T., Yoshida, H., and Yoshino, Y. (2016, January 10–15). Development of a compact Zener DC voltage standard with detachable module system. Proceedings of the 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016), Ottawa, ON, Canada.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Performance evaluation of the ADR1000 low-noise zener IC;2024 Conference on Precision Electromagnetic Measurements (CPEM);2024-07-08
|
|