Abstract
In this paper, we consider the calibration of measuring bridges for non-conventional instrument transformers with digital output. In this context, the main challenge is represented by the necessity of synchronization between analog and digital outputs. To this end, we propose a measurement setup that allows for monitoring and quantifying the main quantities of interest. A possible laboratory implementation is presented and the main sources of uncertainty are discussed. From a metrological point of view, technical specifications and statistical analysis are employed to draw up a rigorous uncertainty budget of the calibration setup. An experimental validation is also provided through the thorough characterization of the measurement accuracy of a commercial device in use at METAS laboratories. The proposed analysis proves how the calibration of measuring bridges for non-conventional instrument transformers requires ad hoc measurement setups and identifies possible space for improvement, particularly in terms of outputs’ synchronization and flexibility of the generation process.
Subject
General Agricultural and Biological Sciences
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献