Retentive Forces and Deformation of Fitting Surface in RPD Clasp Made of Polyether-Ether-Ketone (PEEK)

Author:

Vaddamanu Sunil Kumar1,Alhamoudi Fahad Hussain1ORCID,Chaturvedi Saurabh2ORCID,Alqahtani Nasser M.2,Addas Mohamed Khaled2ORCID,Alfarsi Mohammad Al2,Vyas Rajesh1,Kanji Masroor Ahmed1,Zarbah Mohammad A.2,Alqahtani Waleed M. S.2,Alqahtani Saeed M.2,Abdelmonem Adel M.2,Elmahdi Asim Elsir2

Affiliation:

1. Department of Dental Technology, College of Applied Medical Sciences, King Khalid University, Abha 61421, Saudi Arabia

2. Department of Prosthetic Dentistry, College of Dentistry, King Khalid University, Abha 61421, Saudi Arabia

Abstract

Background: Polyetheretherketone (PEEK) has provided the option to fabricate RPDs with aesthetics unlike metal RPDs, but little attention has been paid to its suitability, especially towards the retentive forces and deformation of the clasp. This study aimed to examine the retentive forces and the fitting surface (inner surface) deformation of clasps made from PEEK and compare it with cobalt–chromium (Co-Cr) clasp. Methods: Forty-two circumferential clasps (14 Co-Cr and 28 PEEK) were fabricated and divided into two groups with clasp undercuts (0.25 mm and 0.5 mm) with thicknesses of 1 mm and 1.5 mm. Each was examined for retentive forces after cycle test on its abutment for 360 cycles. Initial and final retentive forces were recorded. The fitting surface deformation was determined using 3-Matic research analysis software. Results: The results revealed that highest mean initial retentive force was of Co-Cr clasps with 0.50 mm undercut 22.26 N (±10.15 N), and the lowest was the 1 mm PEEK clasps with 0.25 mm undercut 3.35 N (±0.72 N) and highest mean final retentive force was the Co-Cr clasps with 0.50 mm undercut 21.40 N (±9.66 N), and the lowest was the 1 mm PEEK clasps with 0.25 mm undercut 2.71 N (±0.47 N). PEEK clasps had a lower retentive force than Co-Cr clasps with 0.50 undercut. PEEK clasps (1.5 mm) at 0.25 mm undercut had the least deformation (35.3 µm). PEEK showed significantly less deformation (p ≤ 0.014) than Co-Cr. Conclusion: The deformation of PEEK clasps fitting surface was lower than Co-Cr clasps and retentive forces were close to the Co-Cr clasps, suggesting the use of PEEK as an aesthetic clasp option for RPD framework.

Funder

King Khalid University

Publisher

MDPI AG

Subject

Polymers and Plastics,General Chemistry

Reference49 articles.

1. An in vitro comparison of retentive force and deformation of acetal resin and cobalt-chromium clasps;Arda;J. Prosthet. Dent.,2005

2. What is the prevalence of various types of prosthetic dental restorations in Europe?;Zitzmann;Clin. Oral Implant. Res.,2007

3. Effect of acetyl resin retentive arms on the retentive force of circumferential clasps: An in vitro study;Rodrigues;J. Prosthodont. Res.,2012

4. Deflection fatigue of cobalt-chromium, titanium, and gold alloy cast denture clasp;Vallittu;J. Prosthet. Dent.,1995

5. Role of Robotics and Artificial Intelligence in Oral Health and Preventive Dentistry—Knowledge, Perception and Attitude of Dentists;Abouzeid;Oral Health Prev. Dent.,2021