Author:
Bai Lanlan,Gao Yun,Wang Jiajia,Aili Tuerxunnayi,Jia Zhenhong,Lv Xiaoyi,Huang Xiaohui,Yang Jie
Abstract
In this paper, carbon quantum dot-labelled β-lactoglobulin antibodies were used for refractive index magnification, and β-lactoglobulin was detected by angle spectroscopy. In this method, the detection light is provided by a He-Ne laser whose central wavelength is the same as that of the porous silicon microcavity device, and the light source was changed to a parallel beam to illuminate the porous silicon microcavity’ surface by collimating beam expansion, and the reflected light was received on the porous silicon microcavity’ surface by a detector. The angle corresponding to the smallest luminous intensity before and after the onset of immune response was measured by a detector for different concentrations of β-lactoglobulin antigen and carbon quantum dot-labelled β-lactoglobulin antibodies, and the relationship between the variation in angle before and after the immune response was obtained for different concentrations of the β-lactoglobulin antigen. The results of the experiment present that the angle variations changed linearly with increasing β-lactoglobulin antigen concentration before and after the immune response. The limit of detection of β-lactoglobulin by this method was 0.73 μg/L, indicating that the method can be used to detect β-lactoglobulin quickly and conveniently at low cost.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献